Alcohawk PRO Bedienungsanleitung Seite 2

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 2
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 1
If you do not wait 20 minutes before testing, the reading will measure the alcohol in your mouth,
which is many times that related to your blood alcohol level. The detector contains a highly sensitive
semi-conductor sensor, and not waiting the required time could damage (saturate) this sensor and
can cause all future readings to be inaccurate. Please also ensure you do not blow any smoke or
liquid into the unit, or the same damage can result and de-calibrate the detector.
Press the ON/OFF power switch. The detector will start to count down from 200 to 000 on the
display. The warm up and cleaning cycle will prepare the sensor for testing.
After hearing a BEEP and the green READY light comes on, take a deep breath, then gently and
steadily blow into the breath intake until you hear the second BEEP. You will blow for 5 seconds.
Make sure your lips are sealed around the pipe and your hand has not covered the air outlet.
After the red and green display lights blink for 4 seconds, the test results will be displayed by 3
digits and remains for 15 seconds. If “ERR” is displayed, the sample was insufficient (either too
soft, or too short). Press the ON/OFF button again to re-test.
Time out: if you don’t blow within 30 sec from READY, the unit will time out and displays a blinking
OFF”. Press the ON/OFF power switch to turn the unit off and repeat step 1.
How to Analyze the Results?
If the reading is over 050, the red WARN lamp will flash along with a BEEP sound. If you
consume very little alcohol, your actual concentration may be below 5 mg/100ml and the reading
may not be activated. However, the display will read 000
If the sensor is saturated ,the display reads “HOT” and the red lamp will flash along with the
BEEP sound. This can occur if a subject blows alcohol from their saliva into the tester; always leave
at least 20 minutes between drinking and testing. Turn the unit off and leave for 30 minutes before
retesting
If ErH is displayed, the sensor Recognition Technology has detected a contaminated or polluted
sensor and the detector will require re-calibration—please contact us via the details on the back of
this manual for further information.
The UK Drink-Drive limit is 80 mg/100ml (shown as 080 on the AlcoHawk Pro Digital)
Precautions
1. Avoid testing in strong winds, in a closed room with a heavy smoke, or where a lot of
alcohol is being consumed. Do not blow cigarette smoke, food or liquid into the instrument
2. The ALCOHAWK PRO DIGITAL is designed for use in a temperature range of 10 – 40 (C) or
50-104 (F). Operation of the unit in extreme temperature ranges above or below this range
may affect the accuracy of results.
3. Avoid testing in the presence of any substances that contain methyl alcohol, isopropyl
alcohol or acetone. These substances may interfere with the results of the test.
4. Conditions that increase the amount of ketenes on the breath, such as diabetes and low
caloric intake, may cause a false positive test.
5. DO NOT DRINK AND DRIVE. Always have a designated driver if you are drinking
Alcohol.
Always leave at least 20 minutes between drinking and testing
How does it work?
When you drink alcohol it passes through the lining of the stomach into the blood supply. As the
blood moves around your body it is filtered by the liver (which extracts the alcohol over a period of
time) but also passes through the lungs. At the same time as the oxygen in your breath passes
through the alveoli into your blood, a certain amount of the alcohol that is carried by the blood
passes back the other way. It is therefore important to measure deep lung air as this is where the
blood has been in contact with the air in the lungs. A breath from any other part of the lungs will
have a lower alcohol concentration and can give a lower reading than should be the case.
To achieve a consistent reading, a consistent sample of deep lung air is therefore required. Law-
Enforcement devices measure an absolute quantity of air (usually 1.7ltr) each time they are used
and employ a chemical fuel cell to measure alcohol concentration. This however makes such devices
significantly more expensive to manufacture than personal detectors like this that use a “timed blow”
& semi-conductor sensor, but with practice the user can achieve a reasonable level of accuracy by
ensuring that as far as possible the same blow pressure is used each time a sample is taken.
How to get the most from your Breathalyser
When first using the device, 20 minutes after your last drink try testing yourself three times,
approximately two minutes apart, in order to “bracket” the readings. An average reading can then be
assessed, as can the consistency of the blow level required. Note the reading, and then test again
30 minutes later and so on until the unit reads zero.
Try the unit on others; you are likely to find considerable variance in the way in which individuals
metabolise alcohol, depending upon body mass, when they last ate, time of day etc. After a while
you will start to develop a reasonable idea of the way in which you personally react to Alcohol and
you can then modify your drinking appropriately to ensure that you always leave a more than
adequate period after drinking and before driving.
UKBreathalysers would always recommend you do not drive with ANY alcohol in your system, and
that you leave a substantial safety margin between the time you believe you have reached a zero-
level and the point where you may consider driving. Never rely solely upon the readings of this
detector in determining your fitness to drive!
Disposable Mouthpieces
Your detector is equipped with removable / disposable mouthpiece. For all sanitary considerations
replace and dispose the mouthpiece when testing others. To reorder please visit our online store or
call our order desk.
Calibration Procedure
If your device is providing inconsistent, unusually high or low readings, or no readings at all, your
device may need to be re-calibrated. In addition, the breathalyser should be re-calibrated at least
once every 6-12 months. More frequent calibrations may be required depending on frequency of use.
For detailed information on how to have the device re-calibrated, please see our Website.
Cleaning
Although it may seem logical to clean the alcohol detector with alcohol or other cleaning agents, in
reality, this is very harmful to the semiconductor sensor in detector. All semiconductor sensors are
intolerant to high doses (saturation) of alcohol, and other cleaning or anti bacterial agents. Simply
wipe the unit over with a damp (not wet) cloth and dry thoroughly.
AlcoHawk Pro Digital
User Manual
Seitenansicht 1
1 2

Kommentare zu diesen Handbüchern

Keine Kommentare